Wir verwenden Cookies, um für diese Website Funktionen wie z.B. Login, einen Warenkorb oder die Sprachwahl zu ermöglichen. Weiterhin nutzen wir für anonymisierte, statistische Auswertungen der Nutzung Google Analytics, welches Cookies setzt. Mehr Informationen finden Sie in unserer Datenschutzerklärung.
OK, ich bin einverstanden Ich möchte keine Google Analytics-Cookies
The International Journal of Oral & Maxillofacial Implants
Login:
Benutzername:

Kennwort:

Plattform:

Kennwort vergessen?

Registrieren

Int J Oral Maxillofac Implants 19 (2004), Nr. 5     15. Sep. 2004
Int J Oral Maxillofac Implants 19 (2004), Nr. 5  (15.09.2004)

Seite 743-748, Sprache: Englisch


Accuracy of a Manual Torque Application Device for Morse-Taper Implants: A Technical Note
Cehreli, Murat Cavit / Akca, Kývanc / Tönük, Ergin
Purpose: The objective of this study was to compare torques applied by new and used manual torque devices for Morse-taper implants. Material and Methods: Fifteen ITI manual torque devices were tested. Those in group 1 (n = 5) were new (ie, never used), those in group 2 (n = 5) had been used 50 to 200 times, and those in group 3 (n = 5) had been used 500 to 1,000 times. The torques applied by each device were measured for 35 Ncm and 15 Ncm targets in an experimental setup by a custom-made wrench with strain gauges connected to a data acquisition system. The strain-gauge signals were simultaneously delivered to a computer at a sample rate of 10,000 Hz and converted to torque units.
Results: New devices applied higher torques than used devices for the 35-Ncm torque target (P < .05). The torques applied by group 3 devices were approximately 1.5 Ncm lower than those of other groups for the 35-Ncm target and approximately 1 Ncm lower for the 15-Ncm target. Discussion and
Conclusion: ITI manual torque devices deliver consistent torque output, although a slight decrease occurs as a consequence of clinical use.