Application of a Wireless Resonance Frequency Transducer to Assess Primary Stability of Orthodontic Mini-implants: An In Vitro Study in Pig Ilia
Su, Yu-Yu / Wilmes, Benedict / Hönscheid, Ralf / Drescher, Dieter
Purpose: The aim of the study was to evaluate whether resonance frequency analysis (RFA) using a wireless transducer can be used to assess the primary stability of orthodontic mini-implants.
Materials and Methods: Fifteen orthodontic mini-implants were placed in three ilium bone segments of country pigs. The wireless resonance frequency transducer was bonded to the head of the mini-implants, and RFA values of the mini-implants in bone were detected and converted into implant stability quotient (ISQ) values by the RFA monitor. In addition, the percussion test value, peri-implant radiographic bone density, and cortical bone thickness were measured.
Results: The ISQ values of mini-implants correlated linearly with peri-implant radiographic bone density (r = 0.92, P < .0001), cortical bone thickness (r = 0.90, P < .0001), and percussion test values (r = -0.91, P < .0001), respectively. In addition, by means of the calculation of 99% confidence intervals, the absolute values of the three correlation coefficients ranged from 0.61 to 0.98.
Conclusion: This in vitro animal study showed that the presented RFA method using a wireless transducer might have potential to provide an alternative noninvasive assessment of the primary stability of an orthodontic mini-implant.

Keywords: biomechanics, bone screws, dental implants, orthodontic appliance, torque