We are using cookies to implement functions like login, shopping cart or language selection for this website. Furthermore we use Google Analytics to create anonymized statistical reports of the usage which creates Cookies too. You will find more information in our privacy policy.
OK, I agree I do not want Google Analytics-Cookies
The International Journal of Oral & Maxillofacial Implants
Login:
username:

password:

Plattform:

Forgotten password?

Registration

Int J Oral Maxillofac Implants 14 (1999), No. 4     15. Aug. 1999
Int J Oral Maxillofac Implants 14 (1999), No. 4  (15.08.1999)

Page 491-495


Comparison of Interface Relationships Between Implant Components for Laser-welded Titanium Frameworks and Standard Cast Frameworks
Rubenstein, Jeffrey E. / Ma, Tsun
With the introduction of new techniques for the fabrication of frameworks for implant-supported prostheses comes the need to understand how the components used compare to those used for conventional cast frameworks. The relationship of measured machining tolerances between conventional implant components and those components used for stereo laser-welded implant frameworks was determined using a standardized protocol. Statistically significant differences in the horizontal interface relationship were found between paired implant components, which had a mean range from 23.1 to 51.7 µm. From a laboratory and clinical perspective, machining tolerances of implant components represent a variable associated with their manufacturing, which can ultimately affect the fit of a completed prosthesis.

Keywords: implant components, machining tolerances, titanium feet