Int J Oral Maxillofac Implants 3 (1988), No. 1     1. Mar. 1988
Int J Oral Maxillofac Implants 3 (1988), No. 1  (01.03.1988)

Page 57-58


In Vivo Fracture of the Tricalcium Phosphate Coating From the Titanium Body of an Osseointegrating-Type Dental Implant: A Case Report
Edge, Marion J.
The tricalcium phosphate coating on a titanium dental implant fractured in vivo during restorative procedures. Two identical implants were placed at the same visit; one achieved osseointegration and the other failed when, during operator manipulation, the titanium body separated leaving its coating material integrated with the bone.

Keywords: implant, osseointegration, phosphate, titanium, tricalcium