Int J Oral Maxillofac Implants 11 (1996), No. 2     1. Mar. 1996
Int J Oral Maxillofac Implants 11 (1996), No. 2  (01.03.1996)

Page 216-222


Accuracy of Implant Impression Techniques
Assif, David / Marshak, Barry / Schmidt, Avinoam
Three impression techniques were assessed for accuracy in a laboratory cast that simulated clinical practice. The first technique used autopolymerizing acrylic resin to splint the transfer copings. The second involved splinting of the transfer copings directly to an acrylic resin custom tray. In the third, only impression material was used to orient the transfer copings. The accuracy of stone casts with implant analogs was measured against a master framework. The fit of the framework on the casts was tested using strain gauges. The technique using acrylic resin to splint transfer copings in the impression material was significantly more accurate than the two other techniques. Stresses observed in the framework are described and discussed with suggestions to improve clinical and laboratory techniques.

Keywords: abutment, autopolymerizing acrylic resin, implant, transfer copings